Description
The Stat Fax 4700 is ideal for reading ELISA microwell strips in laboratories and institutions where small runs of ELISA assays are routine. Assays can easily be programmed using the touch screen interface with filters ranging from 340 nm – 630 nm. The robust instrument is factory calibrated and maintenance free.
Stat Fax 4700 Semi-automated ELISA Programmable Strip Reader Specification
. Linear Measurement Range: 0.0 to 3.0 Absorbance units (A)
Photometric Accuracy: +/- (1% +/- 0.010) 0.0 through 1.5 ABS
+/- (2% +/- 0.010) 1.5 through 3.0 ABS
Stability: Drift of no more than 0.005A in 8 hours/bichromatic
Light Source: Halogen lamp with lamp saver feature
Standard Wavelengths: Standard: 405, 450, 492, and 630 nm
6-VIS: 405 – 630 nm
6-UV: 340 – 630 nm
Vessel:
Non break apart or break apart strips up to 12 wells long 3 strip loading capacity, (for break apart wells, partial strips or even a single well can be read)
Transport mechanism: Auto-Track™ automatic strip transport Detection Mode
Display: Interactive touch screen 3.5” (8 cm) LCD, color graphic display
Printer: Thermal with graphic capability, approximately 29 characters per line
Speed: Reads, calculates and prints results, in approximately 30 seconds per 12-well strip
Calculation Modes: Single point calibration by standard or factor, multi-point calibration with point-to-point curve fit, linear regressions with log, linear, and log-logit selections.
Test Storage: 120 open channels to store tests Stores all parameters including wavelengths, calculations, unit codes, ranges, interpretations, calibrator values, test names, and previous curve
Dimensions & weight: Approximately 9”x13.5”x5” (24x34x13 cm) weight: 5 lbs (2.3 kg)